Analysis and Testing of a Suitable Compatible Electrode’s Material for Continuous Measurement of Glucose Concentration

机译:用于连续测量葡萄糖浓度的合适兼容电极材料的分析和测试

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The subject of the submitted work is the proposal of electrodes for the continual measurement of the glucose concentration for the purpose of specifying further hemodynamic parameters. The proposal includes the design of the electronic measuring system, the construction of the electrodes themselves and the functionality of the entire system, verified experimentally using various electrode materials. The proposed circuit works on the basis of micro-ammeter measuring the size of the flowing electric current and the electrochemical measurement method is used for specifying the glucose concentration. The electrode system is comprised of two electrodes embedded in a silicon tube. The solution consists of the measurement with three types of materials, which are verified by using three solutions with a precisely given concentration of glucose in the form of a mixed solution and enzyme glucose oxidase. For the testing of the proposed circuit and the selection of a suitable material, the testing did not take place on measurements in whole blood. For the construction of the electrodes, the three most frequently used materials for the construction of electrodes used in clinical practice for sensing biopotentials, specifically the materials Ag/AgCl, Cu and Au, were used. The performed experiments showed that the material Ag/AgCl, which had the greatest sensitivity for the measurement even without the enzyme, was the most suitable material for the electrode. This conclusion is supported by the performed statistical analysis. On the basis of the testing, we can come to the conclusion that even if the Ag/AgCl electrode appears to be the most suitable, showing high stability, gold-plated electrodes showed stability throughout the measurement similarly to Ag/AgCl electrodes, but did not achieve the same qualities in sensitivity and readability of the measured results.
机译:提交的工作的主题是用于持续测量葡萄糖浓度的电极的提议,以指定进一步的血液动力学参数。该提案包括电子测量系统的设计,电极本身的构造和整个系统的功能,使用各种电极材料进行实验验证。所提出的电路基于测量流动电流的尺寸和电化学测量方法的尺寸来指定葡萄糖浓度的基础上。电极系统由嵌入硅管中的两个电极组成。该溶液由具有三种材料的测量组成,通过使用三种溶液来验证,所述三种溶液以混合溶液和酶葡萄糖氧化酶的形式精确地给出的葡萄糖浓度。为了测试所提出的电路和选择合适的材料,测试没有在全血中进行测量。为了对电极的构造,使用用于临床实践中使用的电极构建的三种最常用的材料,用于感测生物能态,特别是材料Ag / AgCl,Cu和Au。所进行的实验表明,即使没有酶的测量敏感性最大的材料Ag / AgCl是最合适的电极的材料。该结论得到了执行的统计分析。在测试的基础上,我们可以得出结论,即使Ag / AgCl电极似乎是最合适的,表示高稳定性,与Ag / AgCl电极相似,镀金电极在整个测量中显示稳定性,但是没有达到相同的敏感性品质和测量结果的可读性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号