首页> 美国卫生研究院文献>International Journal of Particle Therapy >Monte Carlo Simulations of Neutron Ambient Dose Equivalent in a Novel Proton Therapy Facility Design
【2h】

Monte Carlo Simulations of Neutron Ambient Dose Equivalent in a Novel Proton Therapy Facility Design

机译:新型质子治疗设施设计中的中子环境剂量当量的蒙特卡罗模拟

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The neutron shielding properties of the concrete structures of a proposed proton therapy facility were evaluated with help of the Monte Carlo technique. The planned facility's design omits the typical maze-structured entrances to the treatment rooms to facilitate more efficient access and, instead, proposes the use of massive concrete/steel doors. Furthermore, straight conduits in the treatment room walls were used in the design of the facility, necessitating a detailed investigation of the neutron radiation outside the rooms to determine if the design can be applied without violating existing radiation protection regulations. This study was performed to investigate whether the operation of a proton therapy unit using such a facility design will be in compliance with radiation protection requirements.
机译:借助蒙特卡洛技术评估了拟议的质子治疗设施混凝土结构的中子屏蔽性能。计划中的设施的设计省去了通往治疗室的典型迷宫式入口,以便于更有效地进入,而建议使用大型混凝土/钢制门。此外,在设施设计中使用了治疗室壁上的直导管,因此有必要对房间外的中子辐射进行详细调查,以确定是否可以在不违反现有辐射防护规定的情况下进行设计。进行这项研究是为了调查使用这种设施设计的质子治疗仪的操作是否符合辐射防护要求。