Comparison of ANA endpoint dilution using the PolyTiter™ immunofluorescent titration system and conventional serial dilution

机译:使用PolyTiter™免疫荧光滴定系统与常规系列稀释进行ANA终点稀释的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The PolyTiter™ immunofluorescent titration system (PolyTiter system), an attachment for conventional fluorescent microscopes, facilitates immunofluorescent antinuclear antibody (IFA‐ANA) testing by using calibrated light attenuation to determine the semiquantitative endpoint dilution or titer of ANA‐positive specimens. The objective of this study was to compare the determination of ANA titer using the PolyTiter system to that of the conventional serial dilution method. A total of 172 consecutive ANA‐IFA positive serum specimens, including the four most common pattern types, were evaluated in parallel by two readers to determine the dilution endpoint. The two methods produced equal dilution endpoints in 78 specimens (45%), within 1‐dilution endpoint in 121 specimens (70%), and within 2‐dilutions endpoint in 158 specimens (92%). The conventional method consistently produced a more conservative dilution endpoint than the PolyTiter system for three of the four patterns evaluated (i.e., centromere, nucleolar, and speckled vs. homogeneous). The PolyTiter system demonstrated satisfactory performance in comparison to the current standard for determining ANA dilution endpoints, while significantly reducing operator labor, time, and materials usage. J. Clin. Lab. Anal. 16:91–94, 2002. © 2002 Wiley‐Liss, Inc.
机译:PolyTiter™免疫荧光滴定系统(PolyTiter系统)是常规荧光显微镜的附件,它通过使用校准的光衰减来确定ANA阳性样品的半定量终点稀释度或滴定度,促进了免疫荧光抗核抗体(IFA-ANA)测试。这项研究的目的是将使用PolyTiter系统测定的ANA滴度与常规连续稀释法的测定值进行比较。两名读者平行评估了总共172个连续的ANA-IFA阳性血清样本,包括四种最常见的模式类型,以确定稀释终点。两种方法在78个样品(45%)中产生相等的稀释终点,在121个样品中(1-70%)处于1稀释终点内,在158个样品(92%)中在2稀释终点内。对于所评估的四种模式中的三种模式(即着丝粒,核仁和斑点与均质),常规方法始终比PolyTiter系统产生更保守的稀释终点。与目前确定ANA稀释终点的标准相比,PolyTiter系统表现出令人满意的性能,同时大大减少了操作员的劳动,时间和材料使用量。 J.临床实验室肛门16:91–94,2002.©2002 Wiley-Liss,Inc.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号