A recent report indicates that hydroxyzine and its active metabolite cetirizine interfere with the particle‐enhanced turbidimetric inhibition immunoassay (PENTINA) carbamazepine assay. We studied potential interference of hydroxyzine and cetirizine with the turbidimetric carbamazepine immunoassay on ADVIA 1650 and ADVIA 2400 (Bayer Diagnostics, Tarrytown, NY) analyzers. Aliquots of drug‐free serum pools were supplemented with various concentrations of hydroxyzine and cetirizine representing therapeutic, moderate toxic, as well as very toxic concentrations. These samples were assayed by the turbidimetric carbamazepine immunoassay on two analyzers. To study the interference in presence of the analyte, aliquots of a serum pool prepared from patients receiving carbamazepine were further supplemented with various amounts of hydroxyzine and or cetirizine and apparent carbamazepine concentrations were measured again in order to compare with the value of original pool. No apparent carbamazepine concentration was observed when aliquots of drug‐free serum were supplemented with hydroxyzine or cetirizine. Moreover, in the carbamazepine pool, the original carbamazepine concentration compared well when aliquots of this serum pool were further supplemented with hydroxyzine or cetirizine. We conclude that the turbidimetric carbamazepine immunoassay is free from interference of hydroxyzine and cetirizine. J. Clin. Lab. Anal. 21:188–192, 2007. © 2007 Wiley‐Liss, Inc.
机译:最近的一份报告表明,羟嗪及其活性代谢产物西替利嗪干扰了颗粒增强的浊度抑制免疫测定法(PENTINA)卡马西平测定法。我们在ADVIA 1650和ADVIA 2400(Bayer Diagnostics,T​​arrytown,NY)分析仪上用比浊的卡马西平免疫分析法研究了羟嗪和西替利嗪的潜在干扰。在无药血清库的等分试样中补充了各种浓度的羟嗪和西替利嗪,分别代表治疗性,中度毒性和极度毒性。这些样品在两个分析仪上通过比浊卡马西平免疫测定法进行测定。为了研究分析物存在时的干扰,从接受卡马西平的患者制备的血清库的等分试样中进一步补充各种量的羟嗪和/或西替利嗪,并再次测量表观卡马西平浓度,以与原始库的值进行比较。当等份的无药血清补充羟嗪或西替利嗪时,未观察到明显的卡马西平浓度。此外,在卡马西平库中,当进一步向该血清库的等分试样中补充羟嗪或西替利嗪时,原始卡马西平浓度比较好。我们得出的结论是,比浊的卡马西平免疫测定不受羟嗪和西替利嗪的干扰。 J.临床实验室肛门21:188–192,2007。©2007 Wiley-Liss,Inc.

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号