Effect of Tooth Types on the Accuracy of Dental 3D Scanners: An In Vitro Study

机译:牙齿类型对牙科3D扫描仪精度的影​​响:一项体外研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The purpose of this study was to evaluate the accuracy of dental three-dimensional (3D) scanners according to the types of teeth. A computer-aided design (CAD) reference model (CRM) was obtained by scanning the reference typodont model using a high-precision industrial scanner (Solutionix C500, MEDIT). In addition, a CAD test model (CTM) was obtained using seven types of dental 3D scanners (desktop scanners (E1 and DOF Freedom HD) and intraoral scanners (CS3500, CS3600, Trios2, Trios3, and i500)). The 3D inspection software (Geomagic control X, 3DSystems) was used to segment the CRM according to the types of teeth and to superimpose the CTM based on the segmented teeth. The 3D accuracy of the scanner was then analyzed according to the types of teeth. One-way analysis of variance (ANOVA) was used to compare the differences according to the types of teeth in statistical analysis, and the Tukey HSD test was used for post hoc testing (α = 0.05). Both desktop and intraoral scanners showed significant differences in accuracy according to the types of teeth ( < 0.001), and the accuracy of intraoral scanners tended to get worse from anterior to posterior. Therefore, when scanning a complete arch using an intraoral scanner, the clinician should consider the tendency for the accuracy to decrease from anterior to posterior.
机译:这项研究的目的是根据牙齿的类型来评估牙科三维(3D)扫描仪的准确性。通过使用高精度工业扫描仪(Solutionix C500,MEDIT)扫描参考台齿模型,获得了计算机辅助设计(CAD)参考模型(CRM)。此外,使用七种类型的牙科3D扫描仪(台式扫描仪(E1和DOF Freedom HD)和口腔内扫描仪(CS3500,CS3600,Trios2,Trios3和i500))获得了CAD测试模型(CTM)。使用3D检查软件(Geomagic control X,3DSystems)根据牙齿的类型对CRM进行分割,并根据分割后的牙齿叠加CTM。然后根据牙齿的类型分析扫描仪的3D精度。在统计分析中,使用单向方差分析(ANOVA)根据牙齿类型比较差异,并使用Tukey HSD检验进行事后检验(α= 0.05)。台式扫描仪和口腔扫描仪均显示出不同的精确度差异(<0.001),并且口腔扫描仪的精确度从前到后趋于恶化。因此,当使用口内扫描仪扫描完整的牙弓时,临床医生应考虑准确性从前到后降低的趋势。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号