An international collaborative study was performed in order to propose quality control limits for fluconazole disk diffusion susceptibility tests on Mueller-Hinton agar with 2% glucose and 0.5 μg of methylene blue per ml. The supplements may be added before autoclaving the agar, or Mueller-Hinton agar plates may be flooded with a glucose-methylene blue solution. Replicate tests on both types of agar plates generated data to propose zone size limits for tests of Candida parapsilosis ATCC 22019 (22 to 33 mm), C. tropicalis ATCC 750 (26 to 37 mm), and C. albicans ATCC 90028 (28 to 39 mm). C. krusei ATCC 6258 was not useful for this purpose.
机译:进行了一项国际合作研究,以便为每毫升含2%葡萄糖和0.5μg亚甲蓝的Mueller-Hinton琼脂提出氟康唑盘扩散敏感性试验的质量控制限值。可以在高压灭菌琼脂之前添加补充剂,或者可以用葡萄糖-亚甲基蓝溶液淹没Mueller-Hinton琼脂平板。对两种类型的琼脂板进行重复测试,生成数据以提出对假丝酵母ATCC 22019(22至33 mm),热带念珠菌ATCC 750(26至37 mm)和白色念珠菌ATCC 90028(28至28 mm)的区域大小限制39毫米)。 C. krusei ATCC 6258不适用于此目的。