Total hip replacement failure due to adverse local tissue reaction from both ceramic abrasive wear and trunnion corrosion

机译:由于陶瓷磨料磨损和耳轴腐蚀引起的不利的局部组织反应导致完全髋关节置换失败

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The most common sources for metal ions after total hip arthroplasty (THA) are the bearing surface in metal-on-metal articulations and trunnion corrosion. Concomitant dual interface failure is an uncommon complication in metal-on-polyethylene THA. We report an unusual case of a 59-year-old woman with ceramic-on-ceramic THA in 2005, who underwent revision to metal-on-polyethylene THA 4 years later after femoral head fracture. Subsequently, she developed substantial adverse local tissue reaction and significant metal ion elevation and the failure was found to be due to both wear at the bearing surface and corrosion at the head neck junction requiring second revision. Findings included massive adverse local tissue reaction, abductor mechanism destruction, osteolysis, and corrosion damage of the trunnion. Abrasive damage of the trunnion was also noted, but prior abrasion from the original ceramic fracture could not be ruled out. Postoperative course at 14 months demonstrates 95% and 64% reduction in cobalt and chromium levels respectively, with symptom resolution.
机译:全髋关节置换术(THA)之后,最常见的金属离子来源是金属对金属关节和耳轴腐蚀中的轴承表面。伴随的双界面故障是聚乙烯上金属THA的罕见并发症。我们在2005年报告了一位不寻常的病例,一名59岁的女性患有陶瓷-陶瓷THA,在股骨头骨折后4年接受了金属-聚乙烯-THA的修订。随后,她出现了严重的局部组织反应不良和明显的金属离子升高,并且发现这种故障是由于轴承表面的磨损和头部颈部连接处的腐蚀需要再次修正造成的。研究结果包括大量不良的局部组织反应,外展器机制破坏,骨溶解和耳轴的腐蚀损伤。还注意到了耳轴的磨损,但不能排除先前陶瓷断裂造成的先前磨损。术后14个月的病程表明,钴和铬水平分别降低了95%和64%,并且症状得以缓解。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号