Improvements in the processing of polyethylene have led to a dramatic reduction in wear rates in total hip arthroplasty. This led to the adoption of modern highly cross-linked polyethylene in total knee arthroplasty (TKA). However, the differences in modes of wear and failure between total hip arthroplasty and TKA have tempered expectations regarding similar decreases in polyethylene-related complications in TKA. We present a case of early catastrophic failure of a modern sequentially irradiated and annealed highly cross-linked polyethylene insert only 5 years after contemporary cementless TKA.
机译:聚乙烯加工工艺的改进已导致全髋关节置换术的磨损率显着降低。这导致在全膝关节置换术(TKA)中采用了现代高度交联的聚乙烯。然而,全髋关节置换术和TKA之间的磨损和衰竭方式的差异已使人们对TKA中与聚乙烯相关的并发症减少的期望有所改变。我们提出了一种在现代无水泥TKA之后仅5年的现代顺序辐照和退火的高度交联聚乙烯插件发生早期灾难性故障的案例。