摘要

Stepwise-external calibration has previously been shown to produce sub part-per-million (ppm) mass accuracy for the MALDI-FTICR/MS analyses of peptides up to m/z 2500. The present work extends these results to ions up to m/z 4000. Mass measurement errors for ions of higher mass-to-charge are larger than for ions below m/z 2500 when using conventional chirp excitation to detect ions. Mass accuracy obtained by using stored waveform inverse Fourier transform (SWIFT) excitation was evaluated and compared to chirp excitation. Analysis of measurement errors reveals that SWIFT excitation provides smaller deviations from the calibration equation and better mass accuracy than chirp excitation for a wide mass range and for widely varying ion populations.
机译:先前已显示出逐步外部校准可对高达m / z 2500的肽进行MALDI-FTICR / MS分析产生百万分之一(ppm)的质量准确度。目前的工作将这些结果扩展到m / z高达2500的离子。 z4000。当使用常规chi激励检测离子时,荷电更高的离子的质量测量误差大于m / z 2500以下的离子的质量测量误差。评估了通过使用存储的波形逆傅立叶变换(SWIFT)激励获得的质量精度,并将其与线性调频激励进行了比较。对测量误差的分析表明,对于较大的质量范围和变化很大的离子群,SWIFT激发与线性激发相比,提供了与校准方程式较小的偏差,并且具有更高的质量精度。