PurposeOne of the important developments in brachytherapy in recent years has been the clinical implementation of complex modern technical procedures. Today, 3D-imaging has become the standard procedure and it is used for contouring and precise position determination and reconstruction of used brachytherapy applicators. Treatment planning is performed on the basis of these imaging methods, followed by data transfer to the afterloading device. Therefore, checking the entire treatment chain is of high importance. In this work, we describe an end-to-end test for computed tomography (CT)-based brachytherapy with an high-dose-rate (HDR) afterloading device, which fulfills the recommendation of the German radiation-protection-commission.
机译:目的近年来,近距离放射治疗的重要发展之一是复杂现代技术程序的临床实施。如今,3D成像已成为标准程序,可用于轮廓绘制,精确位置确定以及旧近距离放射治疗器的重建。根据这些成像方法执行治疗计划,然后将数据传输到后加载设备。因此,检查整个治疗链非常重要。在这项工作中,我们描述了使用高剂量率(HDR)后装装置的基于计算机断层扫描(CT)的近距离放射疗法的端到端测试,该测试符合德国辐射防护委员会的建议。