Accuracy of insulin pump basal rate delivery, if tested according to the standard IEC 60601-2-24 for infusion pumps, shall be presented as a trumpet curve. This way of graphical presentation is common; however, it is often misunderstood and misinterpreted by people. It is often assumed that a trumpet curve shows the error rate as a function of time, thus implying an increasing accuracy in the course of time. On the contrary, the horizontal axis of a trumpet curve shows increasingly long observation windows. In addition, trumpet curves display only extreme values, that is, those windows with minimal and maximal deviation, which might not be representative for the total deviation. This commentary provides information regarding the calculations and the interpretation of trumpet curves and proposes alternative approaches.
机译:如果按照输液泵的标准IEC 60601-2-24进行测试,则胰岛素泵基础流量的准确度应以喇叭曲线表示。这种图形表示方式很常见。但是,它经常被人们误解和误解。通常认为,喇叭曲线显示的错误率是时间的函数,因此暗示了时间精度的提高。相反,喇叭曲线的水平轴显示的观察窗越来越长。此外,喇叭曲线仅显示极值,即那些具有最小和最大偏差的窗口,这些窗口可能无法代表总偏差。该评论提供了有关小号曲线的计算和解释的信息,并提出了替代方法。