摘要

Glucose meters often have similar performance when compared by error grid analysis. This is one reason that other statistics such as mean absolute relative deviation (MARD) are used to further differentiate performance. The problem with MARD is that too much information is lost. But additional information is available within the A zone of an error grid by using the Taguchi loss function. Applying the Taguchi loss function gives each glucose meter difference from reference a value ranging from 0 (no error) to 1 (error reaches the A zone limit). Values are averaged over all data which provides an indication of risk of an incorrect medical decision. This allows one to differentiate glucose meter performance for the common case where meters have a high percentage of values in the A zone and no values beyond the B zone. Examples are provided using simulated data.
机译:与误差网格分析相比,血糖仪通常具有相似的性能。这是使用其他统计数据(例如平均绝对相对偏差(MARD))进一步区分性能的原因之一。 MARD的问题是丢失了太多信息。但是,通过使用Taguchi损失函数,可以在错误网格的A区域内获得其他信息。应用Taguchi损失函数可使每个血糖仪与参考之间的差值介于0(无错误)到1(错误达到A区域限制)的范围内。该值是所有数据的平均值,可表明存在做出错误医疗决定的风险。这使人们能够区分血糖仪在A区中的百分比值高而B区中没有值的普通情况下的血糖仪性能。使用模拟数据提供了示例。