Background Current Elimination in Thin Layer Ion-Selective Membrane Coulometry

机译:背景电流消除薄层离子选择性膜库仑法

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A promising method for the elimination of undesired capacitive currents in view of realizing a potentially calibration free coulometric ion detection system is presented. The coulometric cell is composed of a porous polypropylene tube doped with a liquid calcium-selective membrane and a silver/silver chloride wire as an inner electrode, forming a thin layer sample between wire and tubing. The total charge passed through the system during potential controlled electrolysis of the thin layer sample is indeed found to be proportional to the amount of calcium present, but non-Faradaic processes do contribute to the obtained signal. We introduce here a multi-pulse procedure that allows one to perform a second excitation pulse at the same excitation potential after exhaustive ion transfer voltammetry of calcium has taken place. The intercept of the calibration curve after background subtraction is found as 20.6 ± 0.6 >μC, significantly lower than the value of 54.1 ± 0.8 >μC for the uncorrected curve. Changes in sample temperature (from 23 °C to 38 °C) did equally not affect the background corrected coulometric readings, supporting that the procedure renders the readout principle more robust.