首页> 美国卫生研究院文献>Journal of Digital Imaging >Accuracy of an Automatic Patient-Positioning System Based on the Correlation of Two Edge Images in Radiotherapy
【2h】

Accuracy of an Automatic Patient-Positioning System Based on the Correlation of Two Edge Images in Radiotherapy

机译:基于放射治疗中两个边缘图像的相关性的自动患者定位系统的精度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We have clinically evaluated the accuracy of an automatic patient-positioning system based on the image correlation of two edge images in radiotherapy. Ninety-six head & neck images from eight patients undergoing proton therapy were compared with a digitally reconstructed radiograph (DRR) of planning CT. Two edge images, a reference image and a test image, were extracted by applying a Canny edge detector algorithm to a DRR and a 2D X-ray image, respectively, of each patient before positioning. In a simulation using a humanoid phantom, performed to verify the effectiveness of the proposed method, no registration errors were observed for given ranges of rotation, pitch, and translation in the >x, >y, and z directions. For real patients, however, there were discrepancies between the automatic positioning method and manual positioning by physicians or technicians. Using edged head coronal- and sagittal-view images, the average differences in registration between these two methods for the >x, >y, and >z directions were 0.11 cm, 0.09 cm and 0.11 cm, respectively, whereas the maximum discrepancies were 0.34 cm, 0.38 cm, and 0.50 cm, respectively. For rotation and pitch, the average registration errors were 0.95° and 1.00°, respectively, and the maximum errors were 3.6° and 2.3°, respectively. The proposed automatic patient-positioning system based on edge image comparison was relatively accurate for head and neck patients. However, image deformation during treatment may render the automatic method less accurate, since the test image many differ significantly from the reference image.
机译:我们已经根据放射治疗中两个边缘图像的图像相关性在临床上评估了自动患者定位系统的准确性。将来自八名接受质子治疗的患者的96幅头部和颈部图像与计划CT的数字重建X射线照片(DRR)进行了比较。通过在定位之前分别对每个患者的DRR和2D X射线图像应用Canny边缘检测器算法来提取两个边缘图像,即参考图像和测试图像。在使用人形模型的仿真中,验证了所提出方法的有效性,在> x ,> y 和z方向。但是,对于实际患者,自动定位方法与医师或技术人员的手动定位之间存在差异。使用边缘冠状冠状和矢状面图像,这两种方法在> x ,> y 和> z 方向上的配准平均差异分别为0.11厘米,0.09厘米和0.11厘米,而最大差异分别为0.34厘米,0.38厘米和0.50厘米。对于旋转和俯仰,平均配准误差分别为0.95°和1.00°,最大误差分别为3.6°和2.3°。所提出的基于边缘图像比较的自动患者定位系统对于头部和颈部患者而言相对准确。但是,由于测试图像与参考图像有很大差异,因此在治疗过程中图像变形可能会使自动方法的准确性降低。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号