Characterization of the effective performance of a high-frequency annular-array based imaging system using anechoic-pipe phantoms

机译:使用消声管幻影的高频环形阵列基于成像系统的有效性能的表征

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A resolution integral (RI) method based on anechoic-pipe, tissue-mimicking phantoms was used to compare the detection capabilities of high-frequency imaging systems based on a single-element transducer, a state-of-the-art, 256-element linear array or a 5-element annular array. All transducers had a central frequency of 40 MHz with similar conventionally measured axial and lateral resolutions (about 50 and 85 μm, respectively). Using the RI metric, the annular array achieved the highest performance (RI = 60), followed by the linear array (47) and the single-element transducer (24). Results showed that the RI metric could be used to efficiently quantify the effective transducer performance and compare the image quality of different systems.
机译:使用基于消声管,组织模仿体模的分辨率积分(RI)方法来比较基于单元素换能器,最先进的256元素的高频成像系统的检测能力线性阵列或5元环形阵列。所有换能器的中心频率均为40 MHz,具有类似的常规轴向和横向分辨率(分别约为50和85μm)。使用RI度量,环形阵列实现了最高性能(RI = 60),其次是线性阵列(47)和单元素换能器(24)。结果表明,RI度量可用于有效地量化有效换能器性能并比较不同系统的图像质量。