首页> 美国卫生研究院文献>other >How accurate is image guided radiation therapy (IGRT) delivered with a micro-irradiator?
【2h】

How accurate is image guided radiation therapy (IGRT) delivered with a micro-irradiator?

机译:是图像引导放射治疗(IGRT)递送的准确性如何用微照射?

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

There is significant interest in delivering precisely targeted small-volume radiation treatments, in the pre-clinical setting, to study dose-volume relationships with tumor control and normal tissue damage. In this work we investigate the IGRT targeting accuracy of the XRad225Cx system from Precision x-Ray using high resolution 3D dosimetry techniques. Initial results revealed a significant targeting error of about 2.4mm. This error was reduced to within 0.5mm after the IGRT hardware and software had been recalibrated. The facility for 3D dosimetry was essential to gain a comprehensive understanding of the targeting error in 3D.
机译:在临床前环境中,提供精确靶向的小剂量放射治疗以研究与肿瘤控制和正常组织损伤之间的剂量-体积关系非常感兴趣。在这项工作中,我们使用高分辨率3D剂量测定技术研究了Precision x-Ray的XRad225Cx系统的IGRT瞄准精度。初步结果显示,明显的瞄准误差约为2.4mm。重新校准IGRT硬件和软件后,此错误减小到0.5mm以内。 3D剂量测定的设施对于全面了解3D定位误差至关重要。