Comprasion of ICare Rebound Tonometer and Goldmann Applanation Tonometer in High Myopia

机译:高度近视眼ICare回弹眼压计和Goldmann压平眼压计的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Purpose. To compare intraocular pressure (IOP) measurements with the Goldmann applanation tonometer (GAT) and the ICare rebound tonometer (RBT) in high myopic eyes. Patients and Methods. This randomized prospective study included 40 eyes of 40 patients with high myopia. All patients' central corneal thickness (CCT), anterior chamber depth (ACD), axial length (AXL), keratometry, and refractive measurements were recorded and followed by IOP measurement with RBT and GAT. Results. The average CCT, AXL, and ACD were determined to be 514.65 ± 32 μm, 27.65 ± 2.22 mm, and 3.25 ± 0.51 mm, respectively. Mean K was 43.27 ± 1.4 D and mean spherical equivalent was −11.31 ± 4.30 D. The mean IOP values obtained by RBT and GAT were 17.18 ± 3.72 mmHg and 16.48 ± 3.19 mmHg, respectively. The deviations of RBT readings from corrected GAT values were highly correlated with CCT values (r = 0.588, P = 0.0001). The mean corrected GAT reading was 17.49 ± 3.01 mmHg. Linear regression analysis showed that a CCT change of 10 μm resulted in an RBT reading deviation of 0.57 mmHg. The Bland-Altman scatter-plot and McNemar test showed a clinically good level of agreement between the two tonometers. Conclusion. This study found a good agreement level between the two tonometers in high myopic patients and that RBT measurements are influenced by CCT variations.
机译:目的。为了比较高眼近视眼中眼压(IOP)与高曼压平眼压计(GAT)和ICare回弹眼压计(RBT)的比较。患者和方法。这项随机前瞻性研究包括40位高度近视患者的40只眼睛。记录所有患者的中央角膜厚度(CCT),前房深度(ACD),轴向长度(AXL),角膜曲率法和屈光测量,然后用RBT和GAT进行IOP测量。结果。测定的平均CCT,AXL和ACD分别为514.65±32μm,27.65±2.22 mm和3.25±0.51 mm。平均K为43.27±1.4 D,平均球当量为-11.31±4.30 D,通过RBT和GAT得到的平均IOP值分别为17.18±3.72 mmHg和16.48±3.19 3.mmHg。 RBT读数与修正的GAT值的偏差与CCT值高度相关(r = 0.588,P = 0.0001)。平均校正后的GAT读数为17.49±3.01mmHg。线性回归分析表明,CCT的变化为10µm,RBT读数偏差为0.57µmmHg。 Bland-Altman散点图和McNemar测试表明,两个眼压计之间的临床一致性水平良好。结论。这项研究发现高度近视患者的两个眼压计之间具有良好的一致性,并且RBT测量值受CCT变化的影响。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号