Rotational distortion correction in endoscopic optical coherence tomography based on speckle decorrelation

机译:基于散斑去相关的内窥镜光学相干断层扫描中的旋转畸变校正

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We present a new technique for the correction of nonuniform rotation distortion in catheter-based optical coherence tomography (OCT), based on the statistics of speckle between A-lines using intensity-based dynamic light scattering. This technique does not rely on tissue features and can be performed on single frames of data, thereby enabling real-time image correction. We demonstrate its suitability in a gastrointestinal balloon-catheter OCT system, determining the actual rotational speed with high temporal resolution, and present corrected cross-sectional and en face views showing significant enhancement of image quality.
机译:我们基于基于强度的动态光散射,基于A线之间的斑点统计数据,提出了一种新技术,用于校正基于导管的光学相干断层扫描(OCT)中非均匀旋转畸变。该技术不依赖于组织特征,可以在单帧数据上执行,从而实现实时图像校正。我们证明了其在胃肠球囊导管OCT系统中的适用性,确定了具有高时间分辨率的实际转速,并提出了校正后的横截面图和正面图,显示了图像质量的显着提高。