Stent and Leaflet Stresses in 26mm First Generation Balloon-Expandable Transcatheter Aortic Valve

机译:26mm第一代球囊扩张式经导管主动脉瓣的支架应力和小叶应力

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

ObjectiveTranscatheter aortic valve replacement is established therapy for high-risk and inoperable patients with severe aortic stenosis, but questions remain regarding long-term durability. Valve design impacts durability. Increased leaflet stresses in surgical bioprostheses have been correlated with degeneration; however, transcatheter valve leaflet stresses are unknown. From 2007–2014, a majority of US patients received first-generation balloon-expandable transcatheter valves. Our goal was to determine stent and leaflet stresses in this valve design using finite element analyses.
机译:目的经导管主动脉瓣置换术已成为治疗高风险且无法手术的严重主动脉瓣狭窄的患者的治疗方法,但长期耐用性仍存在疑问。阀门设计会影响耐用性。外科生物假体中的小叶压力增加与变性有关。但是,经导管瓣膜小叶应力尚不清楚。从2007年至2014年,美国大部分患者接受了第一代球囊扩张式经导管瓣膜。我们的目标是使用有限元分析来确定这种瓣膜设计中的支架和小叶应力。