TLD and OSLD dosimetry systems for remote audits of radiotherapy external beam calibration

机译:TLD和OSLD剂量测定系统用于放射治疗外部束校准的远程审核

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The Imaging and Radiation Oncology Core QA Center in Houston (IROC-H) performs remote dosimetry audits of more than 20,000 megavoltage photon and electron beams each year. Both a thermoluminescent dosimeter (TLD-100) and optically stimulated luminescent dosimeter (OSLD; nanoDot) system are commissioned for this task, with the OSLD system being predominant due to the more time-efficient read-out process. The measurement apparatus includes 3 TLD or 2 OSLD in an acrylic mini-phantom, which are irradiated by the institution under reference geometry. Dosimetry systems are calibrated based on the signal-to-dose conversion established with reference dosimeters irradiated in a Co-60 beam, using a reference dose of 300 cGy for TLD and 100 cGy for OSLD. The uncertainty in the dose determination is 1.3% for TLD and 1.6% for OSLD at the one sigma level. This accuracy allows for a tolerance of ±5% to be used.
机译:休斯敦影像与放射肿瘤学核心质量保证中心(IROC-H)每年对超过20,000兆电压的光子和电子束进行远程剂量学审核。热发光剂量计(TLD-100)和光学激发发光剂量计(OSLD; nanoDot)系统均已为此任务进行调试,由于更省时的读出过程,OSLD系统占主导地位。测量设备包括丙烯酸微型模型中的3个TLD或2个OSLD,它们由机构在参考几何形状下进行辐照。基于在Co-60光束中照射的参考剂量计建立的信号剂量转换,对剂量测定系统进行校准,其中TLD的参考剂量为300 cGy,OSLD的参考剂量为100 cGy。在1 sigma浓度下,TLD剂量确定的不确定度为1.3%,OSLD为1.6%。此精度允许使用±5%的公差。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号