Intraocular pressure profile during the modified diurnal tension curve using Goldman applanation tonometry and dynamic contour tonometry

机译:使用Goldman扁平眼压计和动态轮廓眼压计修正的昼夜张力曲线期间的眼压分布

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The aim of this study was to compare the intraocular pressure (IOP) profile during the modified diurnal tension curve (mDTC) using Goldman applanation tonometry (GAT) and dynamic contour tonometry (DCT) in treated glaucomatous eyes. Eligible subjects were submitted to the mDTC using GAT and DCT in this sequence. IOP measurements were performed at 8 a.m., 10 a.m., 2 p.m., and 4 p.m.. Central corneal thickness was measured using ultrasound pachymetry in the morning. Statistical analysis was performed using paired Student’s t test and Bland–Altman plot. The mean difference between DCT and GAT measurements was 0.9 mmHg. The mean ± SD IOP measurements during the mDTC were 19.68 ± 4.68, 17.63 ± 4.44, 17.25 ± 5.41, and 17.32 ± 4.25 mmHg using GAT and 19.97 ± 4.75, 18.79 ± 4.61, 19.53 ± 5.30, and 19.43 ± 5.45 mmHg using DCT. IOP measurements were higher in the morning (8 a.m.) and decreased throughout the day using both tonometers. The difference between IOP measurements using GAT and DCT was smaller in the morning and increased throughout the day. The IOP variability using GAT was higher than using DCT. Corneal biomechanical properties might help explain our findings.
机译:这项研究的目的是使用高曼压平眼压计(GAT)和动态轮廓眼压计(DCT)比较经治疗的青光眼的修改后的昼夜张力曲线(mDTC)期间的眼压(IOP)曲线。符合资格的受试者按此顺序使用GAT和DCT提交给mDTC。在上午8点,上午10点,下午2点和下午4点进行IOP测量。早晨使用超声测厚仪测量角膜中央厚度。使用成对的Student t检验和Bland-Altman图进行统计分析。 DCT和GAT测量之间的平均差为0.9 mmHg。在mDTC期间,使用GAT和19.97±±4.75、17.79±±4.61、19.53±±5.30和19.43±±5.45 mmHg的mDTC期间的平均SD IOP测量值分别为19.68±0.46、17.63±4.44、17.25±5.45和17.32±4.25 mmHg。使用两个眼压计,IOP测量值在早上(上午8点)较高,并在一天内下降。使用GAT和DCT的IOP测量值之间的差异在早上较小,并在一天中增加。使用GAT的IOP变异性高于使用DCT的IOP变异性。角膜的生物力学特性可能有助于解释我们的发现。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号