The use of a radioactive check device containing Strontium 90 was investigated to carry out the redundancy checks of Farmer‐type ionization chambers. It was not possible to meet the recommended tolerance limits of the redundancy checks without taking into account the angular response of the ionization chambers. The ionization chambers exhibited a maximum variation of 1% in the angular response in this check device. After accounting for the angular response the maximum variation of the short‐term repeatability was 0.14% with a standard deviation of ±0.05%. The long‐term constancy checked over a period of nine months was less than ±0.6% for measurements, taking into account the angular response of the ionization chambers. No significant effect of the background radiation levels was observed on the measurements.PACS number(s): 87.53.–j, 87.52.–g
机译:研究了使用含有锶90的放射性检查装置对Farmer型电离室进行冗余检查。如果不考虑电离室的角度响应,就不可能满足冗余检查的建议公差极限。在该检查装置中,电离室的角度响应最大变化为1%。考虑到角度响应后,短期重复性的最大变化为0.14%,标准偏差为±0.05%。考虑到电离室的角度响应,经过9个月的检查,其长期恒定性小于±0.6%。在测量中未观察到背景辐射水平的显着影响。PACS数量:87.53.–j,87.52.–g