Scatter Correction for Heart SPECT images Using TEW method

机译:使用TEW方法对心脏SPECT图像进行散射校正

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Radionuclide imaging has the potential to be used in quantitative analysis of the regional function of organs. However, quantification of SPECT images is degraded by many factors such as Compton photon scattering. This could have a destructive effect on clinical reports so it is important to do scatter correction to get better quality SPECT images. We intended to determine how scatter correction with the TEW method can help physicians who look at heart SPECT images, get better reports. This study used the TEW method for scatter correction, which was proposed by Ogawa et al.,href="#c9" rid="c9" class=" ref"> (9) using the two narrow windows on either side of the photopeak (20% down and 20% up of the photopeak respectively). Injection of radiopharmaceutical 99mTc was used for medical imaging. In the Shariati Hospital, Tehran, we studied a total of 80 patients with heart disease indications (43 men and 37 women) over the ages of 30–80 years. Contrast and sharpness were considerably improved after scatter correction so physicians could look at defects better. In a few cases scatter correction changed heart defect reports to normal. Using TEW, sensitivity and specificity increased from 86% to 94% and from 61% to 84% respectively. This method was simple to use in clinics.PACS numbers: 87.57.uh
机译:放射性核素成像有潜力用于器官区域功能的定量分析。但是,SPECT图像的量化会因诸如康普顿光子散射之类的许多因素而下降。这可能会对临床报告产生破坏性影响,因此进行散射校正以获得更好质量的SPECT图像非常重要。我们打算确定使用TEW方法进行散射校正如何帮助看心脏SPECT图像的医生获得更好的报告。这项研究使用了Ogawa等人提出的TEW方法进行散射校正, href="#c9" rid="c9" class=" ref">(9) 使用光电峰值两侧的两个狭窄窗口(分别降低光电峰值20%和20%)。注射放射性药物 99m Tc用于医学成像。在德黑兰的Shariati医院,我们研究了80例年龄在30-80岁之间的心脏病患者(43例男性和37例女性)。散点校正后,对比度和清晰度得到了明显改善,因此医生可以更好地观察缺陷。在少数情况下,散布校正将心脏缺陷报告更改为正常。使用TEW,敏感性和特异性分别从86%增加到94%和从61%增加到84%。此方法在诊所中易于使用.PACS编号:87.57.uh

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号