摘要

The purpose of this study was to evaluate beam attenuation by treatment couch and build a treatment couch model in TPS to check for beam–couch intersection at the planning stage and deal with beam attenuation by treatment couch in dose calculation. In this study, a standard treatment couch, Siemens ZXT couch, has been incorporated into Pinnacle3 8.0 TPS, based on an existing TPS tool, model‐based segmentation (MBS). This was done by generating the couch's model from contours of the couch, together with the density information. Both the geometric and dosimetric accuracy of the couch model were evaluated. The test of beam–couch intersection prediction showed good agreement between predicted and measured results, and the differences were within 1° gantry rotation. For individual posterior oblique beams, the attenuation by metallic frames and PMMA couch top could reach nearly as high as 60% and 10%, respectively. For several posterior oblique beams (180°, 220°, 235°) that attenuated by the PMMA couch top, the calculated and measured dose distributions were compared. The dose differences at central axis were within 1%, and almost all points agreed with the calculations when the DD and DTA criteria of 3%/3 mm were adopted. The difference between calculated and measured attenuation factors were within 0.5%. This study demonstrates that the couch model created by MBS, which contains geometric and density information of the couch, can be used to detect the beam–couch intersection, and also is able to provide an accurate representation of the couch top attenuation properties in patient dose calculation.PACS numbers: 87.55.D‐, 87.55.Gh, 87.55.km
机译:这项研究的目的是评估治疗床的光束衰减,并在TPS中建立治疗床模型,以在计划阶段检查光束与沙发的交叉点,并在剂量计算中处理治疗床的光束衰减。在这项研究中,标准治疗床Siemens ZXT床已合并到 Pinnacle 3 8.0 TPS,基于在现有的TPS工具上,基于模型的细分(MBS)。这是通过根据沙发的轮廓以及密度信息生成沙发的模型来完成的。评估了沙发床模型的几何精度和剂量学精度。梁与沙发的交点预测测试表明,预测结果与实测结果吻合良好,并且差异在机架旋转1°以内。对于单个后倾斜梁,金属框架和PMMA沙发顶的衰减分别可以分别高达60%和10%。对于被PMMA沙发顶衰减的几个后倾斜光束(180°,220°,235°),比较了计算和测量的剂量分布。中心轴的剂量差异在1%以内,当采用DD和DTA标准为3%/ 3 mm时,几乎所有点都与计算结果一致。计算衰减系数与测量衰减系数之间的差异在0.5%以内。这项研究表明,MBS创建的沙发模型可以包含束流床的几何和密度信息,可用于检测光束与沙发的交点,还可以准确表示患者剂量下的束流床顶部衰减特性。计算PACS编号:87.55.D-,87.55.Gh,87.55.km

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号