Dose distribution changes with shielding disc misalignments and wrong orientations in breast IOERT: a Monte Carlo – GEANT4 and experimental study

机译:剂量分布随屏蔽盘错位和乳房IOERT方向错误而变化:Monte Carlo – GEANT4和实验研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

One of the most relevant risks in breast intraoperative electron radiotherapy (IOERT) is the incorrect positioning of the shielding disc. If such a setup error occurs, the treatment zone could receive a nonuniform dose delivery, and a considerable part of the electron beam could hit — and irradiate — the patient's healthy tissue. However misalignment and tilt angle of the shielding disc can be evaluated, but it is not possible to measure the corresponding in vivo dose distribution. This led us to develop a simulation using the Geant4 Monte Carlo toolkit to study the effects of disc configuration on dose distribution. Some parameters were investigated: the shielding factor (SF), the radiation back scattering factor (BSF), the volume–dose histogram in the treatment zone, and the maximum leakage dose (MLD) in normal tissue. A lateral shift of the disc (in the plane perpendicular to the beam axis) causes a decrease in SF (from 4% for a misalignment of 5 mm to 40% for a misalignment of 40 mm), but no relevant dose variations were found for a tilt angle until 10°. In the same uncorrected disc positions, the BSF shows no significant change. MLD rises to 3.45 Gy for a 14 mm misalignment and 4.60 Gy for 30° tilt angle when the prescribed dose is 21 Gy. The simulation results are compared with the experimental ones, and allow an a posteriori estimation of the dose distribution in the breast target and underlying healthy tissue. This information could help the surgical team choose a more correct clinical setup, and assist in quantifying the degree of success or failure of an IOERT breast treatment.PACS number: 87.53.Jw, 87.55.dk, 87.55.Gh, 87.55.K‐, 87.55.N‐
机译:乳房术中电子放射疗法(IOERT)中最相关的风险之一是屏蔽盘的位置不正确。如果发生这种设置错误,则治疗区域可能会接收到不均匀的剂量输送,并且相当一部分电子束会撞击并照射患者的健康组织。然而,可以评估屏蔽盘的未对准和倾斜角,但是不可能测量相应的体内剂量分布。这使我们开发了使用Geant4 Monte Carlo工具包进行的仿真,以研究椎间盘配置对剂量分布的影响。研究了一些参数:屏蔽因子(SF),辐射反向散射因子(BSF),治疗区域的体积剂量直方图以及正常组织中的最大泄漏剂量(MLD)。椎间盘的横向移动(在垂直于光束轴的平面上)会导致SF降低(从5mm偏移的4%降低到40mm偏移的40%),但是没有发现相关的剂量变化倾斜角度直到10°。在相同的未校正光盘位置中,BSF没有显示出明显变化。当规定剂量为21 Gy时,未对准14 mm时MLD上升至3.45 Gy,对于30°倾斜角上升至4.60 Gy。仿真结果与实验结果进行了比较,可以对乳房靶标和基础健康组织中的剂量分布进行后验估计。这些信息可以帮助外科手术团队选择更正确的临床设置,并帮助量化IOERT乳房治疗的成功或失败程度.PACS编号:87.53.Jw,87.55.dk,87.55.Gh,87.55.K‐, 87.55。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号