The use of radiochromic EBT2 film for the quality assurance and dosimetric verification of 3D conformal radiotherapy using Microtek ScanMaker 9800XL flatbed scanner

机译:使用Microtek ScanMaker 9800XL平板扫描仪将放射变色EBT2膜用于3D保形放射治疗的质量保证和剂量学验证

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Radiochromic and radiographic films are widely used for radiation dosimetry due to the advantage of high spatial resolution and two‐dimensional dose measurement. Different types of scanners, including various models of flatbed scanners, have been used as part of the dosimetry readout procedure. This paper focuses on the characterization of the EBT2 film response in combination with a Microtek ScanMaker 9800XL scanner and the subsequent use in the dosimetric verification of a 3D conformal radiotherapy treatment. The film reproducibility and scanner uniformity of the Microtek ScanMaker 9800XL was studied. A three‐field 3D conformal radiotherapy treatment was planned on an anthropomorphic phantom and EBT2 film measurements were carried out to verify the treatment. The interfilm reproducibility was found to be 0.25%. Over a period of three months, the films darkened by 1%. The scanner reproducibility was ± 2% and a nonuniformity was ±1.9% along the direction perpendicular to the scan direction. EBT2 measurements showed an underdose of 6.2% at high‐dose region compared to TPS predicted dose. This may be due to the inability of the treatment planning system to predict the correct dose distribution in the presence of tissue inhomogeneities and the uncertainty of the scanner reproducibility and uniformity. The use of EBT2 film in conjunction with the axial CT image of the anthropomorphic phantom allows the evaluation of the anatomical location of dose discrepancies between the EBT2 measured dose distribution and TPS predicted dose distribution.PACS number: 87.55.Qr
机译:由于具有高空间分辨率和二维剂量测量的优势,射线致变色和射线照相胶片被广泛用于辐射剂量测定。不同类型的扫描仪,包括各种型号的平板扫描仪,已被用作剂量测定读出程序的一部分。本文着重于结合Microtek ScanMaker 9800XL扫描仪对EBT2胶片响应进行表征,并随后用于3D保形放射治疗剂量学验证中。研究了Microtek ScanMaker 9800XL的胶片再现性和扫描仪均匀性。计划在拟人模型上进行三场3D保形放射治疗,并进行EBT2胶片测量以验证治疗效果。发现膜间再现性为0.25%。在三个月的时间内,薄膜变暗了1%。扫描仪的重现性为±2%,不均匀度为 < mo>± 1.9 沿垂直于扫描方向的方向。 ETP2测量显示,与TPS预测剂量相比,大剂量区域的剂量不足6.2%。这可能是由于治疗计划系统无法在存在组织不均匀性的情况下预测正确的剂量分布以及扫描仪可重复性和均匀性的不确定性。将EBT2胶片与拟人化体模的轴向CT图像结合使用,可以评估EBT2测得的剂量分布与TPS预测的剂量分布之间的剂量差异的解剖位置.PACS编号:87.55.Qr

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号