首页> 美国卫生研究院文献>Journal of Applied Clinical Medical Physics >Computed tomography dosimetry with high‐resolution detectors commonly used in radiotherapy — an energy dependence study
【2h】

Computed tomography dosimetry with high‐resolution detectors commonly used in radiotherapy — an energy dependence study

机译:放射治疗中常用的具有高分辨率探测器的计算机断层摄影剂量学—能量依赖性研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

New methods of dosimetry in computed tomography (CT) X‐ray fields require the use of high‐resolution detectors instead of pencil‐type ionization chambers typically used for CT dose index (CTDI) measurements. This paper presents a study on the suitability of a wide range of ionization chambers, diodes, and a two‐dimensional detector array, used primarily in radiation therapy, for CT and cone‐beam CT dosimetry. Specifically, the energy dependence of these detectors from 50 kVp up to 125 kVp is reported. All measurements were performed in reference to a calibrated diode for use in this energy region. The radiation quality correction factors provided by the manufacturer were used, depending on the measured half‐value layer (HVL) for the particular X‐ray beam. Our study demonstrated the general usability of thimble ionization chambers. These thimble ionization chambers showed a maximum variation in energy response of 5%. Ionization chambers with even smaller sensitive volume, and which exhibit similar variation in energy dependence, can be used if higher spatial resolution is required. Furthermore, the investigated detectors are better suited for dosimetry at CT and CBCT units than conventional large volume or flat detectors, due to their rotational symmetry. Nevertheless, a flat detector can be used for certain measurement tasks, such as the acquisition of percent depth‐dose curves or beam profiles for nonrotating beams, which are important for beam characterization.PACS numbers: 87.57.uq, 87.56.Da, 87.57.Q‐
机译:计算机断层扫描(CT)X射线场中的剂量学新方法要求使用高分辨率检测器,而不是通常用于CT剂量指数(CTDI)测量的笔型电离室。本文介绍了主要用于放射治疗的各种电离室,二极管和二维检测器阵列适用于CT和锥束CT剂量学的研究。具体地,报道了这些检测器从50 kVp到125 kVp的能量依赖性。所有测量均参考该能量区域中使用的校准二极管进行。使用制造商提供的辐射质量校正因子,具体取决于特定X射线束的测量半值层(HVL)。我们的研究证明了顶针电离室的一般可用性。这些顶针电离室显示出5%的能量响应最大变化。如果需要更高的空间分辨率,则可以使用灵敏体积更小的电离室,并且在能量依赖性方面表现出相似的变化。此外,由于其旋转对称性,与常规的大体积或扁平探测器相比,所研究的探测器更适合在CT和CBCT单元进行剂量测定。尽管如此,平板探测器仍可用于某些测量任务,例如获取深度剂量曲线百分比或非旋转光束的光束轮廓,这对于光束表征至关重要.PACS编号:87.57.uq,87.56.Da,87.57。 Q-

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号