Impact of scanning parameters and breathing patterns on image quality and accuracy of tumor motion reconstruction in 4D CBCT: a phantom study

机译:扫描参数和呼吸模式对4D CBCT中图像质量和肿瘤运动重建准确性的影响:幻像研究

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Four‐dimensional, cone‐beam CT (4D CBCT) substantially reduces respiration‐induced motion blurring artifacts in three‐dimension (3D) CBCT. However, the image quality of 4D CBCT is significantly degraded which may affect its accuracy in localizing a mobile tumor for high‐precision, image‐guided radiation therapy (IGRT). The purpose of this study was to investigate the impact of scanning parameters (hereinafter collectively referred to as scanning sequence) and breathing patterns on the image quality and the accuracy of computed tumor trajectory for a commercial 4D CBCT system, in preparation for its clinical implementation. We simulated a series of periodic and aperiodic sinusoidal breathing patterns with a respiratory motion phantom. The aperiodic pattern was created by varying the period or amplitude of individual sinusoidal breathing cycles. 4D CBCT scans of the phantom were acquired with a manufacturer‐supplied scanning sequence (4D‐S‐slow) and two in‐house modified scanning sequences (4D‐M‐slow and 4D‐M‐fast). While 4D‐S‐slow used small field of view (FOV), partial rotation (200°), and no imaging filter, 4D‐M‐slow and 4D‐M‐fast used medium FOV, full rotation, and the F1 filter. The scanning speed was doubled in 4D‐M‐fast (100°/min gantry rotation). The image quality of the 4D CBCT scans was evaluated using contrast‐to‐noise ratio (CNR), signal‐to‐noise ratio (SNR), and motion blurring ratio (MBR). The trajectory of the moving target was reconstructed by registering each phase of the 4D CBCT with a reference CT. The root‐mean‐squared‐error (RMSE) analysis was used to quantify its accuracy. Significant decrease in CNR and SNR from 3D CBCT to 4D CBCT was observed. The 4D‐S‐slow and 4D‐M‐fast scans had comparable image quality, while the 4D‐M‐slow scans had better performance due to doubled projections. Both CNR and SNR decreased slightly as the breathing period increased, while no dependence on the amplitude was observed. The difference of both CNR and SNR between periodic and aperiodic breathing patterns was insignificant (p  0.48). At end‐exhale phases, the motion blurring was negligible for both periodic and aperiodic breathing patterns; at mid‐inhale phase, the motion blurring increased as the period, the amplitude or the amount of cycle‐to‐cycle variation on amplitude increased. Overall, the accuracy of localizing the moving target in 4D CBCT was within 2 mm under all studied cases. No difference in the RMSEs was noticed among the three scanning sequences. The 4D‐M‐fast scans, free of volume truncation artifacts, exhibited comparable image quality and accuracy in tumor motion reconstruction as the 4D‐S‐slow scans with reduced imaging dose (0.60 cGy vs. 0.99 cGy) due to the use of faster gantry rotation and the F1 filter, suggesting its suitability for clinical use.PACS number: 87.55.Qr
机译:二维锥形束CT(4D CBCT)大大减少了三维(3D)CBCT中呼吸引起的运动模糊伪影。但是,4D CBCT的图像质量显着下降,这可能会影响4D CBCT定位可移动的肿瘤以进行高精度的图像引导放射治疗(IGRT)的准确性。这项研究的目的是调查扫描参数(以下统称为扫描序列)和呼吸模式对商用4D CBCT系统的图像质量和计算的肿瘤轨迹准确性的影响,以为其临床实施做准备。我们用呼吸运动模型模拟了一系列周期性和非周期性的正弦呼吸模式。通过改变各个正弦呼吸周期的周期或幅度来创建非周期性模式。使用制造商提供的扫描序列(4D-S-slow)和两个内部修改的扫描序列(4D-M-slow和4D-M-fast)来获取幻像的4D CBCT扫描。 4D-S-慢速使用小视场(FOV),部分旋转(200°)且不使用成像滤镜,而4D-M-slow和4D-M-fast使用中等FOV,完整旋转和F1滤镜。扫描速度以4D-M-fast(机架旋转速度为100°/ min)加倍。使用对比度噪声比(CNR),信噪比(SNR)和运动模糊比(MBR)评估了4D CBCT扫描的图像质量。通过将4D CBCT的每个相位与参考CT配准来重建运动目标的轨迹。均方根误差(RMSE)分析用于量化其准确性。从3D CBCT到4D CBCT,CNR和SNR显着下降。 4D-S-慢扫描和4D-M-快扫描具有可比的图像质量,而4D-M慢扫描由于投影倍增而具有更好的性能。 CNR和SNR随呼吸周期的增加而略有下降,而未观察到对振幅的依赖性。周期性和非周期性呼吸模式之间的CNR和SNR差异均不显着(p> 0.48)。在呼气末期,周期性和非周期性呼吸模式的运动模糊都可以忽略不计。在吸气中期,运动模糊随着周期,振幅或振幅的逐周期变化量增加而增加。总体而言,在所有研究情况下,在4D CBCT中定位移动目标的精度均在2 mm以内。在三个扫描序列之间,未发现RMSE的差异。 4D-M-快速扫描没有体积截断的伪影,在肿瘤运动重建中表现出与4D-S-慢速扫描相当的图像质量和准确性,这归因于使用更快的方式,降低了成像剂量(0.60 cGy与0.99 cGy)。龙门旋转和F1过滤器,表明它适合临床使用.PACS编号:87.55.Qr

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号