Comparison of TG‐43 dosimetric parameters of brachytherapy sources obtained by three different versions of MCNP codes

机译:三种不同版本的MCNP代码获得的近距离放射治疗源TG‐43剂量参数的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Monte Carlo simulations are widely used for calculation of the dosimetric parameters of brachytherapy sources. MCNP4C2, MCNP5, MCNPX, EGS4, EGSnrc, PTRAN, and GEANT4 are among the most commonly used codes in this field. Each of these codes utilizes a cross‐sectional library for the purpose of simulating different elements and materials with complex chemical compositions. The accuracies of the final outcomes of these simulations are very sensitive to the accuracies of the cross‐sectional libraries. Several investigators have shown that inaccuracies of some of the cross section files have led to errors in  125I and  103Pd parameters. The purpose of this study is to compare the dosimetric parameters of sample brachytherapy sources, calculated with three different versions of the MCNP code — MCNP4C, MCNP5, and MCNPX. In these simulations for each source type, the source and phantom geometries, as well as the number of the photons, were kept identical, thus eliminating the possible uncertainties. The results of these investigations indicate that for low‐energy sources such as  125I and  103Pd there are discrepancies in gL(r) values. Discrepancies up to 21.7% and 28% are observed between MCNP4C and other codes at a distance of 6 cm for  103Pd and 10 cm for  125I from the source, respectively. However, for higher energy sources, the discrepancies in gL(r) values are less than 1.1% for  192Ir and less than 1.2% for  137Cs between the three codes.PACS number(s): 87.56.bg
机译:蒙特卡洛模拟被广泛用于近距离放射治疗源剂量参数的计算。 MCNP4C2,MCNP5,MCNPX,EGS4,EGSnrc,PTRAN和GEANT4是该领域中最常用的代码。这些代码中的每一个都利用一个截面库来模拟具有复杂化学成分的不同元素和材料。这些模拟的最终结果的准确性对截面库的准确性非常敏感。一些调查人员表明,某些横截面文件的不正确导致 125 I和 103 Pd参数出错。本研究的目的是比较使用三种不同版本的MCNP代码(MCNP4C,MCNP5和MCNPX)计算的近距离放射治疗样品来源的剂量参数。在每种源类型的这些模拟中,源和幻像的几何形状以及光子的数量保持相同,从而消除了可能的不确定性。这些调查的结果表明,对于低能量来源,例如 125 I和 103 Pd,gL(r)值存在差异。 MCNP4C与其他代码之间的差异分别为 103 Pd 6 cm和 125 I 10 cm分别达到21.7%和28%。但是,对于更高能量的源,这三个代码之间的 192 Ir的gL(r)值小于1.1%,and 137 Cs的小于1.2%。 PACS编号:87.56.bg

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号