Inappropriate implantable cardioverter defibrillator shocks due to atrial far‐field on the tip‐to‐ring channel for lead dislodgement

机译:尖端到环形通道上的心房远场导致导线移位导致植入式心脏复律除颤器电击不当

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

We reported a case of inappropriate implantable cardioverter defibrillator shocks, due to atrial far‐field on the tip‐to‐ring channel of the fast electrical activity during atrial fibrillation, caused by lead dislocation in the right ventricle outflow tract. During these episodes the can‐to‐right ventricle coil signal correctly recorded the ventricular activity. The shock storm stopped when an antitachycardia pacing restored sinus rhythm.
机译:我们报告了由于不正确的植入式心脏复律除颤器电击引起的情况,这是由于房颤期间快速电活动的尖端环环通道上的心房远场,是由于右心室流出道中的铅脱位引起的。在这些发作期间,从右到右的心室线圈信号正确记录了心室活动。当心动过速起搏恢复窦性心律时,休克风暴停止。