Comparison of ocular biometric measurements between a new swept-source optical coherence tomography and a common optical low coherence reflectometry

机译:新型扫频光学相干断层扫描仪和普通光学低相干反射仪之间的眼睛生物特征测量结果比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The purpose of the current study was to compare the measurements between a new optical biometer based on swept-source optical coherence tomography (SS-OCT), the OA-2000 (Tomey, Japan), and an optical biometer based on optical low coherence reflectometry (OLCR), the Lenstar (Haag-Streit, Switzerland). Ninety-nine eyes of 99 healthy subjects were included. The axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth (AD), lens thickness (LT), keratometry (K) readings, including flat K (Kf), steep K (Ks), mean K (Km), astigmatism vectors J0, J45 at diameters of 2.5 and 3.0 mm, and white-to-white diameter (WTW) were measured three times each using both biometer in normal eyes by random sequence. Bland-Altman analysis showed good agreement between the SS-OCT and OLCR devices for AL, AD, ACD, LT, with narrow 95% LoA (−0.05 to 0.07 mm, −0.09 to 0.10 mm, −0.10 to 0.09 mm, and −0.06 to 0.22 mm, respectively), and the P values of ACD were both >0.05. The CCT, Kf, Ks, Km, J0, J45 and WTW values provided by the OA-2000 were in good agreement with the Lenstar, and statistically significant differences were detected for some of them but not clinical differences. The agreement was excellent especially for AL.
机译:本研究的目的是比较基于扫频光源光学相干断层扫描(SS-OCT)的新型光学生物计,OA-2000(日本托米)和基于光学低相干反射法的光学生物计的测量结果。 (OLCR),Lenstar(瑞士哈格-斯特赖特)。包括99名健康受试者的99眼。轴向长度(AL),中央角膜厚度(CCT),前房深度(ACD),房水深度(AD),晶状体厚度(LT),角膜曲率(K)读数,包括平面K(Kf),陡峭K(Ks ),均值K(Km),散光矢量J0,J45的直径分别为2.5和3.0mm,以及白色至白色直径(WTW)在正常眼睛中均使用这两种生物测定仪以随机顺序分别测量了3次。 Bland-Altman分析显示,SS-OCT和OLCR器件在AL,AD,ACD,LT方面具有良好的一致性,且LoA范围狭窄(95%(-0.05至0.07mm,-0.09至0.10mm,-0.10至0.09mm,以及-分别为0.06至0.22?mm),且ACD的P值均> 0.05。 OA-2000提供的CCT,Kf,Ks,Km,J0,J45和WTW值与Lenstar高度吻合,并且其中一些检测到统计学上的显着差异,但未检测到临床差异。该协议对AL而言尤其出色。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号