We propose a multi-pinhole fluorescent x-ray computed tomography (mp-FXCT) technique for preclinical molecular imaging that can provide the complete data necessary to produce 3-D tomographic images during anaesthesia. In this method, multiple projections are simultaneously acquired through a multi-pinhole collimator with a 2-D detector and full-field volumetric beam to accelerate the data acquisition process and enhance the signal-to-noise ratios of the projections. We constructed a 15-pinhole mp-FXCT imaging system at beamline ARNE-7A at KEK and performed preliminary experiments to investigate its imaging properties using physical phantoms and a non-radioactive I imaging agent. The mp-FXCT system could detect an I concentration of 0.038 mg/ml, the minimum required for in-vivo imaging, at a spatial resolution of about 0.3 mm during a data acquisition time of 90 min, which is less than the time for which anaesthesia is effective and suggests that preclinical molecular imaging is feasible with mp-FXCT.
机译:我们提出了一种用于临床前分子成像的多针孔荧光X射线计算机断层扫描(mp-FXCT)技术,该技术可提供在麻醉期间产生3-D断层图像所需的完整数据。在这种方法中,通过带有2-D检测器的多针孔准直仪和全场体积光束同时获取多个投影,以加快数据采集过程并增强投影的信噪比。我们在KEK的光束线ARNE-7A上构建了一个15针孔的mp-FXCT成像系统,并进行了初步实验,以使用物理体模和非放射性I成像剂研究其成像特性。 mp-FXCT系统可以在90 min的数据采集时间内以约0.3mm的空间分辨率检测到0.038μg/ ml的I浓度,这是体内成像所需的最小值。麻醉是有效的,提示使用mp-FXCT进行临床前分子成像是可行的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号