In this study, we designed an oxygen supply and oxygen saturation monitoring (OSOSM) system. This OSOSM system can provide a continuous supply of oxygen and monitor the peripheral capillary oxygen saturation (SpO2) of patients who accept radiotherapy and use an active breathing coordinator (ABC). A clinical test with 27 volunteers was conducted. The volunteers were divided into two groups based on the tendency of SpO2 decline in breath-holding without the OSOSM system: group A (12 cases) showed a decline in SpO2 of less than 2%, whereas the decline in SpO2 in group B (15 cases) was greater than 2% and reached up to 6% in some cases. The SpO2 of most volunteers declined during rest. The breath-holding time of group A without the OSOSM system was significantly longer than that of group B (p < 0.05) and was extended with the OSOSM system by 26.6% and 27.85% in groups A and B, respectively. The SpO2 recovery time was reduced by 36.1%, and the total rest time was reduced by 27.6% for all volunteers using the OSOSM system. In summary, SpO2 declines during breath-holding and rest time cannot be ignored while applying an ABC. This OSOSM system offers a simple and effective way to monitor SpO2 variation and overcome SpO2 decline, thereby lengthening breath-holding time and shortening rest time.
机译:在这项研究中,我们设计了一个供氧和氧饱和度监测(OSOSM)系统。该OSOSM系统可以提供持续的氧气供应,并监视接受放射疗法并使用主动呼吸协调器(ABC)的患者的外周毛细血管血氧饱和度(SpO2)。对27名志愿者进行了临床测试。根据在没有OSOSM系统的情况下屏气的SpO2下降趋势,将志愿者分为两组:A组(12例)显示SpO2下降小于2%,而B组SpO2下降(15)。大于2%,在某些情况下高达6%。大多数志愿者的SpO2在休息期间都会下降。在没有OSOSM系统的情况下,A组的屏气时间明显长于B组(p <0.05),而在OSOSM系统下,A和B组的屏气时间分别延长了26.6%和27.85%。使用OSOSM系统的所有志愿者的SpO2恢复时间减少了36.1%,总休息时间减少了27.6%。总之,在屏气过程中SpO2下降,而使用ABC时不能忽略休息时间。该OSOSM系统提供了一种简单有效的方法来监控SpO2的变化并克服SpO2的下降,从而延长了屏气时间并缩短了休息时间。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号