摘要

PurposeMapping of epithelial thickness (ET) is useful for detection of keratoconus, a disease characterized by corneal thinning and bulging in which epithelial thinning occurs over the apex. In prior clinical studies, optical coherence tomography (OCT) measurements of ET were systematically thinner than those obtained by 40-MHz high-frequency ultrasound (HFU) where a constant speed of sound (c) of 1636 m/s was used for all corneal layers. The purpose of this work was to study the acoustic properties, that is, c, acoustic impedance (Z), and attenuation (α) of the corneal epithelium and stroma independently using a scanning acoustic microscope (SAM) to investigate the discrepancy between OCT and HFU estimates of ET.
机译:目的上皮厚度(ET)的映射可用于检测圆锥角膜,圆锥角膜是一种以角膜变薄和隆起为特征的疾病,其中角膜上皮发生变薄。在先前的临床研究中,ET的光学相干断层扫描(OCT)测量系统地比通过40 MHz高频超声(HFU)获得的测量更薄,在40 MHz高频超声中,所有角膜均使用恒定的声速(c)1636 m / s层。这项工作的目的是使用扫描声显微镜(SAM)独立研究角膜上皮和基质的声学特性,即c,声阻抗(Z)和衰减(α),以研究OCT与超声之间的差异。 HFU对ET的估计。