>Purpose: Post-therapy quantitative 90Y bremsstrahlung single photon emission computed tomography (SPECT) has shown great potential to provide reliable activity estimates, which are essential for dose verification. Typically 90Y imaging is performed with high- or medium-energy collimators. However, the energy spectrum of 90Y bremsstrahlung photons is substantially different than typical for these collimators. In addition, dosimetry requires quantitative images, and collimators are not typically optimized for such tasks. Optimizing a collimator for 90Y imaging is both novel and potentially important. Conventional optimization methods are not appropriate for 90Y bremsstrahlung photons, which have a continuous and broad energy distribution. In this work, the authors developed a parallel-hole collimator optimization method for quantitative tasks that is particularly applicable to radionuclides with complex emission energy spectra. The authors applied the proposed method to develop an optimal collimator for quantitative 90Y bremsstrahlung SPECT in the context of microsphere radioembolization.>Methods: To account for the effects of the collimator on both the bias and the variance of the activity estimates, the authors used the root mean squared error (RMSE) of the volume of interest activity estimates as the figure of merit (FOM). In the FOM, the bias due to the null space of the image formation process was taken in account. The RMSE was weighted by the inverse mass to reflect the application to dosimetry; for a different application, more relevant weighting could easily be adopted. The authors proposed a parameterization for the collimator that facilitates the incorporation of the important factors (geometric sensitivity, geometric resolution, and septal penetration fraction) determining collimator performance, while keeping the number of free parameters describing the collimator small (i.e., two parameters). To make the optimization results for quantitative 90Y bremsstrahlung SPECT more general, the authors simulated multiple tumors of various sizes in the liver. The authors realistically simulated human anatomy using a digital phantom and the image formation process using a previously validated and computationally efficient method for modeling the image-degrading effects including object scatter, attenuation, and the full collimator-detector response (CDR). The scatter kernels and CDR function tables used in the modeling method were generated using a previously validated Monte Carlo simulation code.>Results: The hole length, hole diameter, and septal thickness of the obtained optimal collimator were 84, 3.5, and 1.4 mm, respectively. Compared to a commercial high-energy general-purpose collimator, the optimal collimator improved the resolution and FOM by 27% and 18%, respectively.>Conclusions: The proposed collimator optimization method may be useful for improving quantitative SPECT imaging for radionuclides with complex energy spectra. The obtained optimal collimator provided a substantial improvement in quantitative performance for the microsphere radioembolization task considered.
机译:>目的:治疗后定量 90 Y ms致辐射单光子发射计算机断层扫描(SPECT)已显示出提供可靠活动估计的巨大潜力,这对于剂量验证至关重要。通常, 90 Y成像是使用高能或中能准直器进行的。但是, 90 Y致stra光子的能谱与这些准直器的典型能谱有很大不同。另外,剂量测定需要定量图像,并且准直仪通常不会针对此类任务进行优化。为 90 Y成像优化准直器既新颖又重要。常规的优化方法不适用于具有连续且宽泛的能量分布的 90 Y stra致辐射光子。在这项工作中,作者开发了一种用于定量任务的平行孔准直器优化方法,该方法特别适用于具有复杂发射能谱的放射性核素。作者应用该方法开发了一种在微球体放射栓塞环境中定量 90 Y ms致辐射SPECT的最佳准直器。>方法:要说明准直器对两种方法的影响由于活动估算的偏差和方差,作者使用了兴趣活动估算量的均方根误差(RMSE)作为品质因数(FOM)。在FOM中,考虑了由于图像形成过程的空白空间引起的偏差。 RMSE由反质量加权,以反映剂量学的应用;对于不同的应用,可以轻松采用更相关的权重。作者提出了准直器的参数化方法,该参数化方法有助于确定准直器性能的重要因素(几何灵敏度,几何分辨率和隔垫渗透率)的纳入,同时保持描述准直器的自由参数的数量较小(即两个参数)。为了使定量的 90 Y ms致病SPECT的优化结果更通用,作者在肝脏中模拟了多种大小的肿瘤。作者使用数字体模逼真地模拟了人体解剖结构,并使用了先前经过验证的高效计算方法对图像形成过程进行了建模,以对包括物体散射,衰减和整个准直仪-检测器响应(CDR)在内的图像降解效果进行建模。建模方法中使用的散射核和CDR函数表是使用先前验证的蒙特卡洛模拟代码生成的。>结果:获得的最佳准直器的孔径,孔径和间隔厚度为84,分别为3.5和1.4毫米。与商用高能通用准直仪相比,最优准直仪分别将分辨率和FOM提高了27%和18%。>结论:提出的准直仪优化方法可能对改善定量SPECT有用对具有复杂能谱的放射性核素进行成像。对于考虑的微球放射性栓塞任务,获得的最佳准直器在定量性能方面提供了实质性的改进。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号