首页> 美国卫生研究院文献>Medical Physics >Phase and sensitivity of receiver coils in magnetic resonance imaging
【2h】

Phase and sensitivity of receiver coils in magnetic resonance imaging

机译:磁共振成像中接收线圈的相位和灵敏度

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Receiver coil response is a major cause of nonuniformities in magnetic resonance images. The spatial dependence of the sensitivity and phase of single-saddle receiver coils has been investigated quantitatively by calculating the >H1 field and comparing the results with measurements of a uniform phantom. Agreement between the measurements and calculations is excellent. A method is developed which corrects for both the nonuniform sensitivity and the phase shifts introduced by receiver coils.
机译:接收器线圈响应是磁共振图像不均匀的主要原因。通过计算> H 1场并将结果与​​均匀幻像的测量结果进行比较,已定量研究了单鞍形接收线圈的灵敏度和相位的空间依赖性。测量和计算之间的一致性非常好。开发了一种方法,该方法既可以校正非均匀灵敏度,又可以校正接收器线圈引入的相移。