The impact of cine EPID image acquisition frame rate on markerless soft-tissue tracking

机译:电影EPID图像采集帧速率对无标记软组织跟踪的影响

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

>Purpose: Although reduction of the cine electronic portal imaging device (EPID) acquisition frame rate through multiple frame averaging may reduce hardware memory burden and decrease image noise, it can hinder the continuity of soft-tissue motion leading to poor autotracking results. The impact of motion blurring and image noise on the tracking performance was investigated.>Methods: Phantom and patient images were acquired at a frame rate of 12.87 Hz with an amorphous silicon portal imager (AS1000, Varian Medical Systems, Palo Alto, CA). The maximum frame rate of 12.87 Hz is imposed by the EPID. Low frame rate images were obtained by continuous frame averaging. A previously validated tracking algorithm was employed for autotracking. The difference between the programmed and autotracked positions of a Las Vegas phantom moving in the superior-inferior direction defined the tracking error (δ). Motion blurring was assessed by measuring the area change of the circle with the greatest depth. Additionally, lung tumors on 1747 frames acquired at 11 field angles from four radiotherapy patients are manually and automatically tracked with varying frame averaging. δ was defined by the position difference of the two tracking methods. Image noise was defined as the standard deviation of the background intensity. Motion blurring and image noise are correlated with δ using Pearson correlation coefficient (R).>Results: For both phantom and patient studies, the autotracking errors increased at frame rates lower than 4.29 Hz. Above 4.29 Hz, changes in errors were negligible with δ < 1.60 mm. Motion blurring and image noise were observed to increase and decrease with frame averaging, respectively. Motion blurring and tracking errors were significantly correlated for the phantom (R = 0.94) and patient studies (R = 0.72). Moderate to poor correlation was found between image noise and tracking error with R −0.58 and −0.19 for both studies, respectively.>Conclusions: Cine EPID image acquisition at the frame rate of at least 4.29 Hz is recommended. Motion blurring in the images with frame rates below 4.29 Hz can significantly reduce the accuracy of autotracking.
机译:>目的:尽管通过多次帧平均降低电影电子门禁成像设备(EPID)采集帧速率可能会减少硬件存储负担并降低图像噪声,但它可能会阻碍软组织运动的持续性,从而导致自动跟踪效果不佳。研究了运动模糊和图像噪声对跟踪性能的影响。>方法:使用非晶硅门禁成像仪(AS1000,Varian Medical Systems,加利福尼亚州帕洛阿尔托)。 EPID规定了12.87 Hz的最大帧速率。通过连续帧平均获得低帧率图像。使用先前验证的跟踪算法进行自动跟踪。沿上下方向移动的拉斯维加斯模型的编程位置和自动跟踪位置之间的差异定义了跟踪误差(δ)。通过测量最大深度的圆的面积变化来评估运动模糊。此外,手动和自动以不同的帧平均跟踪在11个视场角处从4个放射治疗患者获取的1747帧的肺肿瘤。 δ由两种跟踪方法的位置差定义。图像噪声定义为背景强度的标准偏差。使用Pearson相关系数(R),将运动模糊和图像噪声与δ关联。>结果:对于幻像和患者研究,自动跟踪误差均在低于4.29 Hz的帧频下增加。高于4.29 Hz时,δ<1.60 mm时的误差变化可忽略不计。观察到运动模糊和图像噪声分别随帧平均而增加和减少。幻影(R = 0.94)和患者研究(R = 0.72)的运动模糊和跟踪误差显着相关。两项研究的图像噪声和跟踪误差之间均存在中等至较差的相关性。>结论:建议以至少4.29 Hz的帧频采集Cine EPID图像。帧频低于4.29 Hz的图像中的运动模糊会大大降低自动跟踪的准确性。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号