首页> 美国卫生研究院文献>Medical Physics >Technical Note: Defining cyclotron‐based clinical scanning proton machines in a FLUKA Monte Carlo system
【2h】

Technical Note: Defining cyclotron‐based clinical scanning proton machines in a FLUKA Monte Carlo system

机译:技术说明:在FLUKA蒙特卡洛系统中定义基于回旋加速器的临床扫描质子机

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。
获取外文期刊封面目录资料

摘要

PurposeCyclotron‐based pencil beam scanning (PBS) proton machines represent nowadays the majority and most affordable choice for proton therapy facilities, however, their representation in Monte Carlo (MC) codes is more complex than passively scattered proton system‐ or synchrotron‐based PBS machines. This is because degraders are used to decrease the energy from the cyclotron maximum energy to the desired energy, resulting in a unique spot size, divergence, and energy spread depending on the amount of degradation. This manuscript outlines a generalized methodology to characterize a cyclotron‐based PBS machine in a general‐purpose MC code. The code can then be used to generate clinically relevant plans starting from commercial TPS plans.
机译:用途如今,基于回旋加速器的笔形束扫描(PBS)质子机器代表了质子治疗设施的大多数和最负担得起的选择,但是,它们在蒙特卡洛(MC)代码中的表示比被动分散的质子系统或基于同步加速器的PBS机器更为复杂。 。这是因为使用降质剂将回旋加速器的最大能量降低为所需能量,从而导致取决于降解量的独特光斑大小,发散和能量扩散。本手稿概述了通用方法,以通用MC代码来表征基于回旋加速器的PBS机器。然后可以使用该代码从商业TPS计划开始生成临床相关计划。