In this case report, we want to introduce a successful way of applying non-invasive ventilation (NIV) with a full face mask in patients with high cervical spinal cord injury through a novel alarm system for communication. A 57-year-old man was diagnosed with C3 American Spinal Injury Association impairment scale (AIS) B. We applied NIV for treatment of hypercapnia. Because of mouth opening during sleep, a full face mask was the only way to use NIV. However, he could not take off the mask by himself, and this situation caused great fear. To solve this problem, we designed a novel alarm system. The best intended motion of the patient was neck rotation. Sensing was performed by a balloon sensor placed under the head of the patient. A beep sound was generated whenever the pressure was above the threshold, and more than three consecutive beeps within 3,000 ms created a loud alarm for caregivers.
机译:在本病例报告中,我们想通过一种新型的通讯报警系统,介绍一种在全颈椎高位损伤患者中应用全罩式无创通气(NIV)的成功方法。一名57岁的男性被诊断出患有C3美国脊髓损伤协会损害量表(AIS)B。我们应用NIV治疗高碳酸血症。由于在睡觉时张开嘴,全脸面罩是使用NIV的唯一方法。但是,他不能自己摘下面具,这种情况引起了极大的恐惧。为了解决这个问题,我们设计了一种新颖的报警系统。病人最好的运动是颈部旋转。通过放置在患者头部下方的气球传感器进行感应。每当压力高于阈值时,就会发出蜂鸣声,并且在3,000毫秒内连续发出三声以上的蜂鸣声,为护理人员发出响亮的警报。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号