Ankle blood pressure measured by automatic oscillotonometry: a comparison with Doppler pressure measurements.

机译:通过自动示波法测量的脚踝血压:与多普勒压力测量的比较。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Oscillotonometry using Dinamap machine was investigated for the measurement of ankle and brachial blood pressures in our vascular practice. It was validated by comparison with intra-arterial pressure measured by transducer. Systolic ankle and brachial pressures in 43 patients were compared using the Dinamap and Doppler techniques, and significant correlation was found. Ankle and brachial systolic, diastolic and mean pressures and ankle/brachial pressure indices for all three pressures were measured in a group of 12 normal subjects supine at rest, and after treadmill exercise, and a range of normal values defined. The main limitation of the Dinamap is its failure to measure pressures below 50mmHg. The Dinamap technique is a noninvasive, simple, reproducible and quick method of measuring ankle and brachial pressures in vascular surgical practice.
机译:在我们的血管实践中,研究了使用Dinamap机进行示波法测量脚踝和肱动脉血压的方法。通过与换能器测得的动脉内压进行比较,进行了验证。使用Dinamap和Doppler技术比较了43例患者的收缩期踝关节和肱动脉压,发现了显着相关性。在静息时,跑步机运动后仰卧的一组12名正常受试者的一组中,测量了这三种压力的踝关节和肱动脉收缩压,舒张压和平均压力以及踝关节/肱压力指数,并定义了一系列正常值。 Dinamap的主要局限性在于它无法测量低于50mmHg的压力。 Dinamap技术是一种无创,简单,可重现,快速的血管外科手术中测量踝关节和肱动脉压力的方法。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号