We compared the accuracy and precision of two microbiological methods and one high-pressure liquid chromatography (HPLC) procedure used to measure the concentrations of flucytosine in serum. On the basis of an analysis of six standards, all methods were judged reliable within acceptable limits for clinical use. With the biological methods, a slight loss of linearity was observed in the 75- to 100-micrograms/ml range. Compared with the bioassays, the HPLC method did not present linearity problems and was more precise and accurate in the critical zone of 100 micrograms/ml. On average, results obtained with patient sera containing 50 to 100 micrograms of flucytosine per ml were 10.6% higher with the HPLC method than with the bioassays. Standards for the biological assays may be prepared in serum or water.
机译:我们比较了两种微生物学方法和一种用于测量血清中氟胞嘧啶浓度的高压液相色谱(HPLC)程序的准确性和精密度。在分析六种标准的基础上,所有方法均被认为在临床使用可接受的范围内是可靠的。使用生物学方法,在75至100微克/毫升的范围内观察到线性的轻微损失。与生物测定法相比,HPLC方法没有线性问题,并且在100微克/毫升的临界区中更加精确。平均而言,采用HPLC方法的患者血清每毫升含50至100微克氟胞嘧啶所获得的结果要比采用生物测定方法高出10.6%。可以在血清或水中制备用于生物学测定的标准品。