The electrophoretic mobilities of three bacterial strains were investigated by capillary electrophoresis (CE) and were compared with results obtained by microelectrophoresis (ME). The CE measurements yielded bimodal electropherograms for two of the strains, thus illustrating for the first time that surface charge variations within a monoclonal population can be probed by CE. Intrapopulation variations were not detected by ME. The mobilities of three chemically distinct types of latex microspheres were also measured. Differences between the mean mobilities obtained by CE and ME were not statistically significant (P ≤ 0.50); the standard deviations of the CE measurements were typically 2 to 10 times smaller than those obtained by comparable ME measurements. The reproducibility of CE permitted batch-to-batch mobility variations to be probed for the bacteria (one of the strains exhibited such variations), and aggregation was evident in one of the latex suspensions. These effects were not measurable with ME.
机译:通过毛细管电泳(CE)研究了三种细菌菌株的电泳迁移率,并将其与通过微电泳(ME)获得的结果进行了比较。 CE测量产生了其中两个菌株的双峰电泳图,因此首次说明了CE可以探测单克隆群体内的表面电荷变化。 ME未检测到种群内变异。还测量了三种化学上不同类型的乳胶微球的迁移率。 CE和ME获得的平均迁移率之间的差异无统计学意义(P≤0.50); CE测量的标准偏差通常比可比较的ME测量获得的标准偏差小2至10倍。 CE的重现性允许探查细菌的批次间迁移率变异(一种菌株表现出这种变异),并且在一种胶乳悬浮液中明显聚集。这些效果用ME无法测量。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号