Mass Airflow Cabinet for Control of Airborne Infection of Laboratory Rodents

机译:控制实验室啮齿动物空气传播的大流量柜

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A mass airflow cabinet for handling and housing of laboratory rodents has been developed and tested. The unit consists of a high-efficiency particulate air filter and uniform distribution of air at a vertical velocity of 19 cm per s. Animals are maintained without bedding in mesh-bottomed cages that rest on rollers for rotation inside the cabinet. There is an air barrier of 90 cm per s separating the cabinet air from room air. Sampling for airborne bacteria yielded an average of 0.03 colony-forming units (CFU) per ft3 of air inside the cabinet, whereas 28.8 CFU per ft3 was simultaneously detected outside the cabinet during housekeeping, a reduction of almost three logs. The efficiency of the air barrier was tested by aerosolization of T3 phage. When phage was aerosolized 5 cm outside the cabinet, no phage could be detected 5 cm inside when the fans were operating; with the fans off an average of 1.6 × 104 plaque-forming units (PFU) per ft3 was detected in six tests. Aerosolization of phage inside the cabinet yielded an average of 9 × 10 PFU per ft3 outside; an average of 4.1 × 106 PFU per ft3 were detected with the fans not in operation, a reduction of more than four logs. In-use studies on effectiveness showed that the cabinet significantly reduced the incidence of mice originally titer-free to Reo-3 virus. Hemagglutination inhibition antibodies to Reo-3 were detected in 9/22 (42%) mice housed in a conventionally ventilated animal laboratory while no seroconversion was detected in any of 22 mice housed in the mass air flow cabinet in the same laboratory.
机译:已经开发并测试了用于处理和容纳实验室啮齿动物的大气流柜。该装置由高效的颗粒空气过滤器和以19 cm / s的垂直速度均匀分布的空气组成。饲养动物时,不要将它们放在网状底部的笼子中,该笼子放在滚子上以便在柜子内旋转。橱柜空气与室内空气之间存在每秒90厘米的空气屏障。采样空气传播的细菌后,柜内空气平均每ft 3 产生0.03个菌落形成单位(CFU),而在室外空气中同时检测到每ft 3 28.8个CFU。内务处理期间减少了近三原木。通过雾化T3噬菌体来测试空气屏障的效率。当将噬菌体雾化到机柜外部5 cm处时,当风扇运行时,在5 cm内无法检测到噬菌体;在六个测试中,每扇ft 3 的风扇平均关闭了1.6×10 4 噬菌斑形成单位(PFU)。机柜内噬菌体的气雾化作用产生了外部每ft 3 平均9×10 PFU;在风扇不运行的情况下,每ft 3 平均检测到4.1×10 6 PFU,减少了四个以上的原木。在有效性方面的使用研究表明,该柜子大大降低了最初对Reo-3病毒无滴度的小鼠的发病率。在常规通风的动物实验室中饲养的9/22(42%)小鼠中检测到针对Reo-3的血凝抑制抗体,而在同一实验室的空气流通柜中饲养的22只小鼠中均未检测到血清转化。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号