Clinical evaluation and comparison of the Infrasonde Arteriosonde and mercury sphygmomanometer in measurement of blood pressure in children.

机译:在儿童血压测量中使用InfrasondeArteriosonde和汞血压计进行临床评估和比较。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Both systolic and diastolic pressures can be measured in children with the mercury sphygmomanometer, the Arteriosonde and the Infrasonde. Estimates made blindly with these instruments were compared with directly measured intra-arterial blood pressures in 50 children aged between 4 days and 14 years. Systolic and diastolic estimates with the three techniques showed highly significant correlations with simultaneous intra-arterial measurements (P less than 0.001). The Infrasonde diastolic estimates were least satisfactory and the slope of the regression line against the intra-arterial pressure differed significantly from unity (y=0.54x+29.53). In 11 small children a satisfactory diastolic estimate could not be obtained with the mercury sphygmomanometer. While the mercury sphygmomanometer should remain the standard hospital equipment, an Arteriosonde would be valuable if it is difficult to hear Korotkoff's sounds on auscultation and if a diastolic pressure is required. For research investigations into childhood blood pressure an Arteriosonde or mercury sphygmomanometer, coupled with a device to exclude observer bias, is probably most suitable. Although the Infrasonde is not sufficiently accurate for research purposes, it is acceptable for routine ward use.
机译:可以使用水银血压计,Arteriosonde和Infrasonde测量儿童的收缩压和舒张压。用这些仪器盲目进行的估计与直接测量的50位4天至14岁儿童的动脉内血压进行了比较。三种技术的收缩压和舒张压估计值与同时进行的动脉内测量值显示高度相关性(P小于0.001)。 Infrasonde舒张压的估计最不令人满意,并且回归线相对于动脉内压的斜率与单位差异显着不同(y = 0.54x + 29.53)。在11个小孩中,使用汞血压计无法获得令人满意的舒张压估计值。尽管水银血压计应仍然是医院的标准设备,但如果很难听见科罗特科夫在听诊时的声音并且需要舒张压,则Arteriosonde将是很有价值的。对于儿童血压的研究调查,最合适的方法是使用Arteriosonde或汞血压计,再加上消除观察者偏倚的装置。尽管对于研究目的而言,Infrasonde不够准确,但对于常规病房使用来说还是可以接受的。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号