Assessment of Doppler ultrasound to measure systolic and diastolic blood pressures in infants and young children.

机译:评估多普勒超声以测量婴幼儿的收缩压和舒张压。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

A recently developed instrument uses the Doppler shift technique to detect vessel wall movement, and it has been suggested that in conjunction with a conventional sphygmomanometer systolic and diastolic blood pressures can be measured. A controlled study was carried out in 20 children recovering from cardiac surgery where direct intra-arterial measurements (one observer) were compared with independent measurements using the Doppler instrument (2 observers). Systolic pressures measured directly and by Doppler technique correlated well and there was no significant difference between intra-arterial and indirect measurements whether the latter were taken by doctors or by nurses. In contrast, direct and indirect diastolic pressure measurements correlated poorly and were significantly overestimated with a mean difference of 6-25 mmHg (range +25 to -10) for doctors, and 4-25 mmHg (range +20 to -10) for nurses. Thus, the instrument adequately measured systolic blood pressure, but in our hands did not give precise measurements for diastolic blood pressure.
机译:最近开发的仪器使用多普勒频移技术来检测血管壁的运动,并且已经提出,可以与常规的血压计一起测量收缩压和舒张压。在20名从心脏手术中恢复过来的儿童中进行了一项对照研究,将直接动脉内测量(一名观察员)与使用多普勒仪进行的独立测量(两名观察员)进行了比较。直接和通过多普勒技术测量的收缩压相关性很好,无论是医生还是由护士进行的动脉内测量和间接测量之间都没有显着差异。相比之下,直接和间接舒张压测量值的相关性很差,并且被医生高估了,其平均差异为6-25 mmHg(+25至-10),而护士为4-25 mmHg(+20至-10) 。因此,该仪器足以测量收缩压,但在我们手中却无法准确测量舒张压。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号