Constrained acetabular liners have been developed for patients who are at high risk for dislocation or who are undergoing revision surgery for recurrent dislocations. We report on 2 cases of failure of tripolar constrained liners due to severe backside polyethylene wear after dissociation of the outer polyethylene liner without dislocation, a mode of failure not previously reported. The backside of the inserts suffered severe polyethylene deformation, wear, and scratching due to dissociation from the locking mechanism. In patients with tripolar constrained liners, radiographic evidence of eccentric wear should be considered as possible occult dissociation of the polyethylene liner within the shell. Conversion to a modular dual mobility liner appears to be a viable solution in this setting.
机译:约束性髋臼内衬已为高位错位风险或因复发性位错而接受翻修手术的患者开发。我们报告了2例三极约束衬套失效的情况,这些情况是由于聚乙烯外衬套解离后没有严重脱位而造成的严重的背面聚乙烯磨损,这种失效方式以前没有报道过。由于与锁定机构分离,插入件的背面遭受了严重的聚乙烯变形,磨损和刮擦。在三极约束衬里的患者中,应考虑到偏心磨损的射线照相证据可能是聚乙烯衬里在壳体内的隐匿解离。在这种情况下,转换为模块化双移动衬管似乎是可行的解决方案。