Differential impact of multi-focus fan beam collimation with L-mode and conventional systems on the accuracy of myocardial perfusion imaging: Quantitative evaluation using phantoms

机译:L模式和常规系统的多焦点扇形光束准直对心肌灌注成像准确性的不同影响:使用体模的定量评估

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

Objective(s):A novel IQ-SPECT™ method has become widely used in clinical studies. The present study compares the quality of myocardial perfusion images (MPI) acquired using the IQ-SPECT™ (IQ-mode), conventional (180° apart: C-mode) and L-mode (90° apart: L-mode) systems. We assessed spatial resolution, image reproducibility and quantifiability using various physical phantoms.
机译:目标:一种新颖的IQ-SPECT™方法已在临床研究中广泛使用。本研究比较了使用IQ-SPECT™(IQ模式),常规(相隔180°:C模式)和L模式(相隔90°:L模式)系统获取的心肌灌注图像(MPI)的质量。 。我们使用各种物理模型评估了空间分辨率,图像可再现性和可量化性。