Calibration of a prototype NIRS oximeter against two commercial devices on a blood-lipid phantom

机译:针对血腥体模上的两个商用设备校准NIRS血氧仪原型

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

In a blood-lipid liquid phantom the prototype near-infrared spectroscopy oximeter OxyPrem was calibrated against the INVOS® 5100c adult sensor in respect to values of regional tissue oxygen haemoglobin saturation (rStO2) for possible inclusion in the randomised clinical trial - SafeBoosC. In addition different commercial NIRS oximeters were compared on changing haemoglobin oxygen saturation and compared against co-oximetry. The best calibration was achieved with a simple offset and a linear scaling of the OxyPrem rStO2 values. The INVOS adult and pediatric sensor gave systematically different values, while the difference between the NIRO® 300 and the two INVOS sensors were magnitude dependent. The co-oximetry proved unreliable on such low haemoglobin and high Intralipid levels.
机译:在血脂液体模型中,针对局部组织氧血红蛋白饱和度(rStO2)值,针对INVOS®5100c成人传感器对原型近红外光谱血氧仪OxyPrem进行了校准,以便可能纳入随机临床试验-SafeBoosC。另外,比较了不同的商用NIRS血氧饱和度计在改变血红蛋白氧饱和度方面的作用,并与共血氧测定法进行了比较。通过简单的偏移和OxyPrem rStO2值的线性比例可达到最佳校准。成人和儿童INVOS传感器给出的系统值不同,而NIRO®300和两个INVOS传感器之间的差异取决于幅度。协同血氧测定法在如此低的血红蛋白和高血脂水平上是不可靠的。