Comparison of corneal endothelial cell measurements by two non-contact specular microscopes

机译:两种非接触式镜面显微镜对角膜内皮细胞测量结果的比较

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

BackgroundMeasurement of corneal endothelial cell density is important both for clinical diagnosis as well as clinical studies. Since endothelial cell loss is considered irreversible in humans, even small changes in endothelial cell density are relevant. Therefore it is important to know whether different instruments for endothelial cell density measurements give the same results and can thus be used interchangeably. In this study we compare corneal endothelial cell density and morphometry measurements from two widely used non-contact specular microscopes, the Topcon SP3000P and Konan Noncon Robo SP8000.
机译:背景技术角膜内皮细胞密度的测量对于临床诊断和临床研究都是重要的。由于内皮细胞的丢失在人类中被认为是不可逆的,因此即使内皮细胞密度发生很小的变化也很重要。因此,重要的是要知道用于内皮细胞密度测量的不同仪器是否给出相同的结果,因此可以互换使用。在这项研究中,我们比较了Topcon SP3000P和Konan Noncon Robo SP8000这两种广泛使用的非接触式镜面显微镜对角膜内皮细胞密度和形态学测量的影响。