A comparison of cycloplegic and manifest refractions on the NR-1000F (an objective Auto Refractometer).

机译:NR-1000F(客观自动验光仪)上的睫状肌麻痹和明显折射的比较。

代理获取
本网站仅为用户提供外文OA文献查询和代理获取服务,本网站没有原文。下单后我们将采用程序或人工为您竭诚获取高质量的原文,但由于OA文献来源多样且变更频繁,仍可能出现获取不到、文献不完整或与标题不符等情况,如果获取不到我们将提供退款服务。请知悉。

摘要

The manifest (dry) and cycloplegic refractions of 50 eyes of 25 patients aged 8 to 28 years were studied on the Nikon Auto Refractometer NR-1000F (AR) and compared with the results of clinical refraction (CR) under homatropine and the final clinical acceptance on postmydriatic testing. Only patients in the younger age groups with low to moderate refractive errors were included in this study; high myopes and hypermetropes and patients with aphakia and mixed astigmatism were excluded. The degree of agreement for spherical equivalents, sphere components, and cylinder components was analysed separately for both cycloplegic and manifest refractions on the AR and CR. The results showed that the fixation target in the NR-1000F induces significant instrument myopia during manifest refraction in the younger patients with lower refractive errors. We recommend that caution should be exercised in interpreting manifest refractions on the AR, especially in younger patients. A cycloplegic automatic refraction would be acceptably accurate.
机译:在尼康自动验光仪NR-1000F(AR)上研究了25例8至28岁患者的50只眼的明显(干性)和睫状肌屈光度,并与荷马碱在临床上的屈光度(CR)结果和最终临床接受度进行了比较经过散瞳测试。本研究仅包括低度至中度屈光不正的年轻患者。高度近视和远视及无晶状体和混合散光的患者被排除在外。分别针对AR和CR上的睫状肌麻痹和明显屈光分别分析了球面等效物,球面分量和圆柱分量的一致程度。结果表明,NR-1000F的固定目标在明显屈光度较低的年轻患者中,在明显屈光期间可诱发明显的器械近视。我们建议在解释AR的明显折射时应格外小心,尤其是在年轻患者中。睫状肌麻痹自动验光将是可以接受的准确度。

著录项

相似文献

  • 外文文献
  • 中文文献
  • 专利
代理获取

客服邮箱:kefu@zhangqiaokeyan.com

京公网安备:11010802029741号 ICP备案号:京ICP备15016152号-6 六维联合信息科技 (北京) 有限公司©版权所有
  • 客服微信

  • 服务号